Description
Drager Savina 300 Features
The Drager Savina 300 Ventilator was designed as a turbine-driven ventilation system. With the technology of a Turbine built in the Savina 300 can offer quick presence times and deliver continuous high flow ventilation – up to 250 l/min. The system offers non-invasive ventilation for both adults and pediatric patients. The Savina 300 Ventilator offers a variety of different ventilation modes, volume, and pressure support along with spontaneous breathing support. The system has a highly effective sensitive leak management incorporated into the ventilation system reducing ventilation times. The Drager Savina 300 comes with a large color touchscreen which makes it easier to configure to each individual patient’s needs.
- Intuitive for simple operation and quick configuration
- Draeger-wide standardized user interface provides confidence in use and reduces training time
- Quick operational readiness with an automatic device check
- Intelligent alarm handling for a quick response to patient alarm situations
- Smooth and sealed surfaces for easy cleaning and disinfection
- Huge range of ventilation modes (e.g. PC-APRV, VC-MMV, AutoFlow)
- Stress-free spontaneous breathing with excellent trigger response time thanks to the turbine
- Free breathing with AutoFlow in volume constant ventilation at a minimum pressure level
Drager Savina 300 Specifications
Ventilation Modes
- VC-CMV / VC-AC
- VC-SIMV
- VC-MMV (optional)
- PC-APRV (optional)
- PC-BIPAP1) / PC-SIMV (optional)
- PC-AC (optional)
- SPN-CPAP
Optional Enhancements
- AutoFlow – Automatic adaption of the inspiratory flow in volume orientated ventilation modes
- NIV – Non Invasive Ventilation with optimized alarm systems and automatic leakage compensation
- Capnography – Mainstream CO2 measurement
- MonitoringPlus – Loops, Trends, user Logbook
- LPO – Low Pressure Oxygen. Independant oxygen supply
- Nurse call – Connection for transmitting alarm signals to a central, alarm system
General
- Patient type: Adult, pediatric
- Respiratory rate: 2/min to 80/min
- Inspiration time: 0.2 to 10 s
- Tidal volume: 0.05 to 2.0 L, BTPS2) with option PediatricPlus 0.02 to 2.0 L
- Inspiratory pressure:: 1 to 99 mbar (or hPa or cmH2O)
- PEEP/interm. PEEP 0 to 50 mbar (or hPa or cmH2O)
- Pressure support/ΔPsupp: 0 to 50 mbar (or hPa or cmH2O) (relative to PEEP)
- Flow acceleration: 5 to 200 mbar/s (or hPa/s or cmH2O/s)
- O2-concentration: 21 to 100 Vol. %
- Trigger sensitivity (Flow trigger): 1 to 15 L/min
- Inspiratory termination criterion: 5 to 75 % PIF (peak inspiratory flow)
- PC-APRV (optional):
- Inspiratory time Thigh 0.2 to 22.0 s
- Expiratory time Tlow 0.1 to 22.0 s
- Inspiratory pressure Phigh 1 to 95 mbar (or hPA or cmH2O)
- Expiratory pressure Plow 0 to 50 mbar (or hPA or cmH2O)
Displayed Measured Values
- Airway pressure measurements Max. airway pressure, plateau pressure, mean airway pressure, PEEP 0 to 99 mbar (or hPa or cmH2O)
- Minute volume (MV): Total MV, spontaneous MV 0 to 99 L/min, BTPS
- Tidal volume: Inspiratory VT, expiratory VTe, VTspon 0 to 3999 mL, BTPS
- Total respiratory rate: Total and spontaneous respiratory rate, 0 to 150/min
- Inspiratory O2-concentration: 21 to 100 % Vol.
- End-tidal CO2 concentration: EtCO2 0 to 100 mmHg (or 0 to 13.2 Vol% or 0 to 13.3 kPa)
- Breathing gas temperature: 18 to 48 °C (64.4 to 118.4 °F)
- Curve displays: Paw(t), Flow (t), Tidal volume (t), CO2 (t)
- Ventilation ratio (I:E): 1:150 to 150:1
- Compliance C: 0.5 to 200 mL/mbar (or mL/hPa or mL/cmH2O)
- Resistance R: 3 to 300 mbar/L/s (or hPa/L/s or cmH2O/L/s)
- Leakage minute volume MVleak: 0 to 100 %
- Rapid shallow breathing RSB: 0 to 9999 (1/min/L)
- Special Maneuvers (optional):
- Intrinsic PEEP PEEPi 0 to 100 mbar (or hPa or cmH2O)
- Exp. Hold
Alarms
- Airway pressures: high / low
- Expiratory minute volume: high / low
- Tidal volume: high / low
- Apnea-alarm time: 15 to 60 sec
- Spontaneous breathing frequency: high
- Inspiratory O2-concentration: high / low
- Inspiratory breathing gas temperature: high
- EtCO2: high / low
Performance Data
- Maximum (continuous) inspiratory flow: 250 L/min
- Valve response time T0…90: ≤ 5 ms
- Control principle: time-cycled, volume-controlled, pressure limited
- Safety valve opening pressure: 120 mbar (or hPa or cmH2O)
- Emergency valve: automatically enables spontaneous breathing with filtered ambient air if air and O2 supply should fail.
- Automatic gas switch-over function if O2 supply fails
- Output for pneumatic medication nebulizer: synchronized with inspiration
- Leak compensation:
- optimized patient-ventilator synchrony adjusts the flow trigger and the inspiratory termination criteria for leaks.
- tube application: up to 10L/min
- NIV VC-modes: up to 25 L/min
- NIV PC-modes: unlimited
Operating Data
- Mains power connection: 100 V to 240 V, 50/60 Hz
- Current consumption: max. 1.3 A at 240 V, max. 3.4 A at 100 V
- Battery: internal typically 45 min (optional extension up to 5 h)
- Turbine exchange interval: 8 years, with no limit in operating hours during this interval
Digital Machine Outputs
- Digital output and input via an RS 232 C interface
- Dräger MEDIBUS and MEDIBUS.X
Gas Supply
- Air Turbine technology
- O2 gas supply 3 bar (43.5 psi) – 10 % up to 6 bar (87 psi)
Dimensions and Weights
- Dimensions W x H x D (without trolley) 460 x 383 x 364 ±2 mm (18.11 x 15.08 x 14.33 ±0.08 inch)
- Weight (basic device) approx. 26 kg (57.3 lbs) without trolley
- Diagonal screen size 12″ TFT color touch screen